Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: HP 54602A: (1 year) CAL VER /5520+SC600 DATE: 30-May-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 47 NUMBER OF LINES: 522 CONFIGURATION: Fluke 5520A (S6) ============================================================================= # # Source: Hewlett Packard 54600-Series User and Service Guide, June 1992 # # Compatibility: 5520/CAL or MET/CAL 5.0 or later # # Subprocedures: None # # Required Files: None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N J F W 1.002 ASK+ D 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature : 20 - 30 degC 1.004 DISP Humidity : <80% 1.005 HEAD [32] WARNING! 1.006 DISP 1.006 DISP DURING THE FOLLOWING PROCEDURE, HIGH VOLTAGES THAT COULD 1.006 DISP CAUSE SERIOUS INJURY OR EVEN DEATH ARE PRESENT! 1.006 DISP 1.006 DISP USE CAUTION TO PREVENT INJURY! 1.007 DISP This procedure requires the following... 1.007 DISP 1.007 DISP Digital Multimeter; 0.1% mV resolution, 0.01% accuracy 1.008 HEAD {DC CALIBRATOR} 1.009 DISP Connect the multimeter to the rear panel DC Calibrator. 1.010 DISP Press PRINT/UTILITY 1.010 DISP 1.010 DISP Press the Self Test softkey, then press the DAC softkey. 1.011 HEAD DC CALIBRATOR: {Low Level} 1.012 EVAL Is the {DC Calibrator output 0.00 VDC +/-500 uV}? 2.001 HEAD DC CALIBRATOR: {High Level} 2.002 DISP Press any key to continue the test. 2.003 EVAL Is the {DC Calibrator output 5.00 VDC +/-10 mV}? 3.001 HEAD {VOLTAGE MEASUREMENT ACCURACY} 3.002 RSLT =CH 1 3.003 DISP MAKE THE FOLLOWING CONNECTION... 3.003 DISP 3.003 DISP 5520 NORMAL OUTPUT ----------- CH 1 INPUT 3.004 DISP Press SETUP, then press the Default Setup softkey. 3.004 DISP 3.004 DISP Press VOLTAGE, then press the V avg softkey. 3.005 DISP Set the Volts/Div to 5 V/div. 3.006 DISP Adjust the CH 1 Position knob to place the baseline 3.006 DISP near (but not at) the bottom of the display. 3.007 DISP Press CURSORS, then press the V1 softkey. 3.008 5520 0V S 2W 3.009 DISP Using the cursors knob, set V1 cursor on the baseline. 3.009 DISP 3.009 DISP If you are in an electrically noisy environment, 3.009 DISP it can help to place a shorting cap on the input BNC 3.009 DISP connector when positioning V1. 3.010 HEAD VOLTAGE MEASUREMENT ACCURACY: {5 V/div} 3.011 5520 35V S 2W 3.012 DISP Press the V2 softkey, then position the V2 cursor 3.012 DISP to the baseline. 3.013 EVAL Is {Delta V readout between 34.315 and 35.685 V}? 4.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {2 V/div} 4.002 DISP Set the Volts/Div to 2 V/div. 4.003 5520 0V S 2W 4.004 DISP Using the cursors knob, set V1 cursor on the baseline. 4.004 DISP 4.005 5520 14V S 2W 4.006 DISP Press the V2 softkey, then position the V2 cursor 4.006 DISP to the baseline. 4.007 EVAL Is {Delta V readout between 13.726 and 14.274 V}? 5.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {1 V/div} 5.002 DISP Set the Volts/Div to 1 V/div. 5.003 5520 0V S 2W 5.004 DISP Using the cursors knob, set V1 cursor on the baseline. 5.004 DISP 5.005 5520 7V S 2W 5.006 DISP Press the V2 softkey, then position the V2 cursor 5.006 DISP to the baseline. 5.007 EVAL Is {Delta V readout between 6.863 and 7.137 V}? 6.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {0.5 V/div} 6.002 DISP Set the Volts/Div to 0.5 V/div. 6.003 5520 0V S 2W 6.004 DISP Using the cursors knob, set V1 cursor on the baseline. 6.004 DISP 6.005 5520 3.5V S 2W 6.006 DISP Press the V2 softkey, then position the V2 cursor 6.006 DISP to the baseline. 6.007 EVAL Is {Delta V readout between 3.4315 and 3.5685 V}? 7.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {0.2 V/div} 7.002 DISP Set the Volts/Div to 0.2 V/div. 7.003 5520 0V S 2W 7.004 DISP Using the cursors knob, set V1 cursor on the baseline. 7.004 DISP 7.005 5520 1.4V S 2W 7.006 DISP Press the V2 softkey, then position the V2 cursor 7.006 DISP to the baseline. 7.007 EVAL Is {Delta V readout between 1.3726 and 1.4274 V}? 8.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {0.1 V/div} 8.002 DISP Set the Volts/Div to 0.1 V/div. 8.003 5520 0V S 2W 8.004 DISP Using the cursors knob, set V1 cursor on the baseline. 8.004 DISP 8.005 5520 700mV S 2W 8.006 DISP Press the V2 softkey, then position the V2 cursor 8.006 DISP to the baseline. 8.007 EVAL Is {Delta V readout between 686.3 and 713.7 mV}? 9.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {50 mV/div} 9.002 DISP Set the Volts/Div to 50 mV/div. 9.003 5520 0V S 2W 9.004 DISP Using the cursors knob, set V1 cursor on the baseline. 9.004 DISP 9.005 5520 350mV S 2W 9.006 DISP Press the V2 softkey, then position the V2 cursor 9.006 DISP to the baseline. 9.007 EVAL Is {Delta V readout between 343.15 and 356.85 mV}? 10.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {20 mV/div} 10.002 DISP Set the Volts/Div to 20 mV/div. 10.003 5520 0V S 2W 10.004 DISP Using the cursors knob, set V1 cursor on the baseline. 10.004 DISP 10.005 5520 140mV S 2W 10.006 DISP Press the V2 softkey, then position the V2 cursor 10.006 DISP to the baseline. 10.007 EVAL Is {Delta V readout between 137.26 and 142.74 mV}? 11.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {10 mV/div} 11.002 DISP Set the Volts/Div to 10 mV/div. 11.003 5520 0V S 2W 11.004 DISP Using the cursors knob, set V1 cursor on the baseline. 11.004 DISP 11.005 5520 70mV S 2W 11.006 DISP Press the V2 softkey, then position the V2 cursor 11.006 DISP to the baseline. 11.007 EVAL Is {Delta V readout between 68.63 and 71.37 mV}? 12.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {5 mV/div} 12.002 DISP Set the Volts/Div to 5 mV/div. 12.003 5520 0V S 2W 12.004 DISP Using the cursors knob, set V1 cursor on the baseline. 12.004 DISP 12.005 5520 35mV S 2W 12.006 DISP Press the V2 softkey, then position the V2 cursor 12.006 DISP to the baseline. 12.007 EVAL Is {Delta V readout between 34.155 and 35.845 mV}? 13.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {2 mV/div} 13.002 DISP Set the Volts/Div to 2 mV/div. 13.003 5520 0V S 2W 13.004 DISP Using the cursors knob, set V1 cursor on the baseline. 13.004 DISP 13.005 5520 14mV S 2W 13.006 DISP Press the V2 softkey, then position the V2 cursor 13.006 DISP to the baseline. 13.007 EVAL Is {Delta V readout between 13.47 and 14.53 mV}? 14.001 RSLT =CH 2 14.002 DISP MAKE THE FOLLOWING CONNECTION... 14.002 DISP 14.002 DISP 5520 NORMAL OUTPUT ----------- CH 2 INPUT 14.003 5520 1.0V S 2W 14.004 DISP Press AUTOSCALE 14.004 DISP 14.005 5520 * S 14.006 DISP Press VOLTAGE, then press the V avg softkey. 14.007 DISP Set the Volts/Div to 5 V/div. 14.008 DISP Adjust the CH 2 Position knob to place the baseline 14.008 DISP near (but not at) the bottom of the display. 14.009 5520 0V S 2W 14.010 DISP Press CURSORS, then press the V1 softkey. 14.010 DISP Using the cursors knob, set V1 cursor on the baseline. 14.011 HEAD VOLTAGE MEASUREMENT ACCURACY: {5 V/div} 14.012 5520 35V S 2W 14.013 DISP Press the V2 softkey, then position the V2 cursor 14.013 DISP to the baseline. 14.014 EVAL Is {Delta V readout between 34.315 and 35.685 V}? 15.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {2 V/div} 15.002 DISP Set the Volts/Div to 2 V/div. 15.003 5520 0V S 2W 15.004 DISP Using the cursors knob, set V1 cursor on the baseline. 15.004 DISP 15.005 5520 14V S 2W 15.006 DISP Press the V2 softkey, then position the V2 cursor 15.006 DISP to the baseline. 15.007 EVAL Is {Delta V readout between 13.726 and 14.274 V}? 16.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {1 V/div} 16.002 DISP Set the Volts/Div to 1 V/div. 16.003 5520 0V S 2W 16.004 DISP Using the cursors knob, set V1 cursor on the baseline. 16.004 DISP 16.005 5520 7V S 2W 16.006 DISP Press the V2 softkey, then position the V2 cursor 16.006 DISP to the baseline. 16.007 EVAL Is {Delta V readout between 6.863 and 7.137 V}? 17.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {0.5 V/div} 17.002 DISP Set the Volts/Div to 0.5 V/div. 17.003 5520 0V S 2W 17.004 DISP Using the cursors knob, set V1 cursor on the baseline. 17.004 DISP 17.005 5520 3.5V S 2W 17.006 DISP Press the V2 softkey, then position the V2 cursor 17.006 DISP to the baseline. 17.007 EVAL Is {Delta V readout between 3.4315 and 3.5685 V}? 18.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {0.2 V/div} 18.002 DISP Set the Volts/Div to 0.2 V/div. 18.003 5520 0V S 2W 18.004 DISP Using the cursors knob, set V1 cursor on the baseline. 18.004 DISP 18.005 5520 1.4V S 2W 18.006 DISP Press the V2 softkey, then position the V2 cursor 18.006 DISP to the baseline. 18.007 EVAL Is {Delta V readout between 1.3726 and 1.4274 V}? 19.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {0.1 V/div} 19.002 DISP Set the Volts/Div to 0.1 V/div. 19.003 5520 0V S 2W 19.004 DISP Using the cursors knob, set V1 cursor on the baseline. 19.004 DISP 19.005 5520 700mV S 2W 19.006 DISP Press the V2 softkey, then position the V2 cursor 19.006 DISP to the baseline. 19.007 EVAL Is {Delta V readout between 686.3 and 713.7 mV}? 20.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {50 mV/div} 20.002 DISP Set the Volts/Div to 50 mV/div. 20.003 5520 0V S 2W 20.004 DISP Using the cursors knob, set V1 cursor on the baseline. 20.004 DISP 20.005 5520 350mV S 2W 20.006 DISP Press the V2 softkey, then position the V2 cursor 20.006 DISP to the baseline. 20.007 EVAL Is {Delta V readout between 343.15 and 356.85 mV}? 21.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {20 mV/div} 21.002 DISP Set the Volts/Div to 20 mV/div. 21.003 5520 0V S 2W 21.004 DISP Using the cursors knob, set V1 cursor on the baseline. 21.004 DISP 21.005 5520 140mV S 2W 21.006 DISP Press the V2 softkey, then position the V2 cursor 21.006 DISP to the baseline. 21.007 EVAL Is {Delta V readout between 137.26 and 142.74 mV}? 22.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {10 mV/div} 22.002 DISP Set the Volts/Div to 10 mV/div. 22.003 5520 0V S 2W 22.004 DISP Using the cursors knob, set V1 cursor on the baseline. 22.004 DISP 22.005 5520 70mV S 2W 22.006 DISP Press the V2 softkey, then position the V2 cursor 22.006 DISP to the baseline. 22.007 EVAL Is {Delta V readout between 68.63 and 71.37 mV}? 23.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {5 mV/div} 23.002 DISP Set the Volts/Div to 5 mV/div. 23.003 5520 0V S 2W 23.004 DISP Using the cursors knob, set V1 cursor on the baseline. 23.004 DISP 23.005 5520 35mV S 2W 23.006 DISP Press the V2 softkey, then position the V2 cursor 23.006 DISP to the baseline. 23.007 EVAL Is {Delta V readout between 34.155 and 35.845 mV}? 24.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {2 mV/div} 24.002 DISP Set the Volts/Div to 2 mV/div. 24.003 5520 0V S 2W 24.004 DISP Using the cursors knob, set V1 cursor on the baseline. 24.004 DISP 24.005 5520 14mV S 2W 24.006 DISP Press the V2 softkey, then position the V2 cursor 24.006 DISP to the baseline. 24.007 EVAL Is {Delta V readout between 13.47 and 14.53 mV}? 25.001 RSLT =CH 3 25.002 DISP MAKE THE FOLLOWING CONNECTION... 25.002 DISP 25.002 DISP 5520 NORMAL OUTPUT ----------- CH 3 INPUT 25.003 5520 1V 1kH SI S 2W 25.004 DISP Press AUTOSCALE 25.004 DISP 25.005 5520 * S 25.006 DISP Press VOLTAGE, then press the V avg softkey. 25.007 DISP Set the Volts/Div to .5 V/div. 25.008 DISP Adjust the CH 3 Position knob to place the baseline 25.008 DISP near (but not at) the bottom of the display. 25.009 5520 0V S 2W 25.010 DISP Press CURSORS, then press the V1 softkey. 25.010 DISP Using the cursors knob, set V1 cursor on the baseline. 25.011 HEAD VOLTAGE MEASUREMENT ACCURACY: {0.5 V/div} 25.012 5520 3.5V S 2W 25.013 DISP Press the V2 softkey, then position the V2 cursor 25.013 DISP to the baseline. 25.014 EVAL Is {Delta V readout between 3.4315 and 3.5685 V}? 26.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {0.1 V/div} 26.002 DISP Set the Volts/Div to 0.1 V/div. 26.003 5520 0V S 2W 26.004 DISP Using the cursors knob, set V1 cursor on the baseline. 26.004 DISP 26.005 5520 700mV S 2W 26.006 DISP Press the V2 softkey, then position the V2 cursor 26.006 DISP to the baseline. 26.007 EVAL Is {Delta V readout between 686.3 and 713.7 mV}? 27.001 RSLT =CH 4 27.002 DISP MAKE THE FOLLOWING CONNECTION... 27.002 DISP 27.002 DISP 5520 NORMAL OUTPUT ----------- CH 4 INPUT 27.003 5520 1.0V 1kH SI S 2W 27.004 DISP Press AUTOSCALE 27.004 DISP 27.005 5520 * S 27.006 DISP Press VOLTAGE, then press the V avg softkey. 27.007 DISP Set the Volts/Div to .5 V/div. 27.008 DISP Adjust the CH 4 Position knob to place the baseline 27.008 DISP near (but not at) the bottom of the display. 27.009 DISP Press CURSORS, then press the V1 softkey. 27.010 5520 0V S 2W 27.011 DISP Using the cursors knob, set V1 cursor on the baseline. 27.012 HEAD VOLTAGE MEASUREMENT ACCURACY: {0.5 V/div} 27.013 5520 3.5V S 2W 27.014 DISP Press the V2 softkey, then position the V2 cursor 27.014 DISP to the baseline. 27.015 EVAL Is {Delta V readout between 3.4315 and 3.5685 V}? 28.001 HEAD VOLTAGE MEASUREMENT ACCURACY: {0.1 V/div} 28.002 DISP Set the Volts/Div to 0.1 V/div. 28.003 5520 0V S 2W 28.004 DISP Using the cursors knob, set V1 cursor on the baseline. 28.004 DISP 28.005 5520 700mV S 2W 28.006 DISP Press the V2 softkey, then position the V2 cursor 28.006 DISP to the baseline. 28.007 EVAL Is {Delta V readout between 686.3 and 713.7 mV}? 29.001 HEAD {BANDWIDTH} 29.002 RSLT =CH 1 29.003 DISP MAKE THE FOLLOWING CONNECTION... 29.003 DISP 29.003 DISP 5520 SCOPE OUTPUT ----------50 Ohm--CH 1 INPUT 29.004 5520 800mVpp 250kH LS S6 S L 29.005 DISP Press AUTOSCALE. 29.006 MESS Slew the amplitude for 8 divisions. 29.007 5520 800mVpp 250kH LS S6 N L 29.008 DISP Change the Sweep Speed to 5 nS. 29.008 DISP 29.009 MESS Slew the frequency for 5.66 divisions. 29.010 5520 150MH -1000% mVpp LS S6 L 30.001 RSLT =CH 2 30.002 DISP MAKE THE FOLLOWING CONNECTION... 30.002 DISP 30.002 DISP 5520 SCOPE OUTPUT ----------50 Ohm--CH 2 INPUT 30.003 5520 800mVpp 250kH LS S6 S L 30.004 DISP Press AUTOSCALE. 30.005 MESS Slew the amplitude for 8 divisions. 30.006 5520 800mVpp 250kH LS S6 N L 30.007 DISP Change the Sweep Speed to 5 nS. 30.007 DISP 30.008 MESS Slew the frequency for 5.66 divisions. 30.009 5520 150MH -1000% mVpp LS S6 L 31.001 RSLT =CH 3 31.002 DISP MAKE THE FOLLOWING CONNECTION... 31.002 DISP 31.002 DISP 5520 SCOPE OUTPUT ----------50 Ohm--CH 3 INPUT 31.003 5520 800mVpp 250kH LS S6 S L 31.004 DISP Press AUTOSCALE. 31.005 DISP Set the Volts/Div to 0.1 V/div. 31.006 MESS Slew the amplitude for 8 divisions. 31.007 5520 800mVpp 250kH LS S6 N L 31.008 MESS 31.009 DISP Change the Sweep Speed to 5 nS. 31.009 DISP 31.010 5520 250MH mVpp LS S6 S L 31.011 EVAL Is {the amplitude greater than 5.66 divisions}? 32.001 MESS 32.002 RSLT =CH 4 32.003 DISP MAKE THE FOLLOWING CONNECTION... 32.003 DISP 32.003 DISP 5520 SCOPE OUTPUT ----------50 Ohm--CH 4 INPUT 32.004 5520 800mVpp 250kH LS S6 S L 32.005 DISP Press AUTOSCALE. 32.006 DISP Set the Volts/Div to 0.1 V/div. 32.007 MESS Slew the amplitude for 8 divisions. 32.008 5520 800mVpp 250kH LS S6 N L 32.009 MESS 32.010 DISP Change the Sweep Speed to 5 nS. 32.010 DISP 32.011 5520 250MH mVpp LS S6 S L 32.012 EVAL Is {the amplitude greater than 5.66 divisions}? 33.001 MESS 33.002 ASK+ U 33.003 HEAD {HORIZONTAL DELTA T AND 1/DELTA T ACCURACY} 33.004 DISP MAKE THE FOLLOWING CONNECTION... 33.004 DISP 33.004 DISP 5520 SCOPE OUTPUT ----------50 Ohm--CH 1 INPUT 33.005 HEAD HORIZONTAL DELTA T AND 1/DELTA T ACCURACY: {20 uS/div} 33.006 5520 0.1mT M1 S6 S L 33.007 DISP Press SETUP, then press the Default Setup softkey. 33.007 DISP 33.007 DISP Press AUTOSCALE. 33.008 DISP Set the time base to 20 uS/div. 33.009 DISP Adjust the trigger level to obtain a stable display. 33.009 DISP 33.010 DISP Press TIME, then press the Freq and Period softkeys. 33.011 EVAL Is {Frequency readout between 9.899 and 10.10 kHz}? 34.001 5520 0.1mT M1 S6 S L 34.002 EVAL Is {Period readout between 98.98 and 101.01 uS}? 35.001 HEAD HORIZONTAL DELTA T AND 1/DELTA T ACCURACY: {200 nS/div} 35.002 5520 1uT M1 S6 S L 35.003 DISP Set the time base to 200 nS/div. 35.004 DISP Adjust the trigger level to obtain a stable display. 35.004 DISP 35.005 EVAL Is {Frequency readout between 989.8 kHz and 1.0104 MHz}? 36.001 5520 1uT M1 S6 S L 36.002 EVAL Is {Period readout between 989.7 ns and 1.010 uS}? 37.001 HEAD HORIZONTAL DELTA T AND 1/DELTA T ACCURACY: {5 nS/div} 37.002 5520 10nT M4 S6 S L 37.003 DISP Set the time base to 5 nS/div. 37.004 DISP Set the Volts/Div to 0.5 V/div. 37.005 DISP Use the Position control to center the signal. 37.006 DISP Adjust the trigger level to obtain a stable display. 37.006 DISP 37.007 EVAL Is {Frequency readout between 97.078 and 103.1 MHz}? 38.001 5520 10nT M4 S6 S L 38.002 EVAL Is {Period readout between 9.699 and 10.30 nS}? 39.001 HEAD {CH 1 INTERNAL TRIGGER SENSITIVITY} 39.002 ASK- U 39.003 DISP Press SETUP, then press the Default Setup softkey. 39.004 HEAD CH 1 TRIGGER SENSITIVITY: {25 MHz @ 0.35 div}? 39.005 5520 50mVpp 25MH LS S6 S L 39.006 DISP Press AUTOSCALE. 39.007 DISP Set the Volts/Div to 0.1 V/div. 39.008 MESS Slew the amplitude for 0.35 divisions. 39.009 5520 50mVpp 25MH LS S6 N L 39.010 MESS Adjust Trigger Level if needed for a triggered display. 39.010 MESS 39.011 EVAL Will the {display trigger}? 40.001 MESS 40.002 HEAD CH 1 TRIGGER SENSITIVITY: {100 MHz @ 1.0 div}? 40.003 5520 100mVpp 150MH LS S6 S L 40.004 DISP Press AUTOSCALE. 40.005 DISP Set the Volts/Div to 0.1 V/div. 40.006 MESS Slew the amplitude for 1.0 division. 40.007 5520 100mVpp 150MH LS S6 N L 40.008 MESS Adjust Trigger Level if needed for a triggered display. 40.008 MESS 40.009 EVAL Will the {display trigger}? 41.001 MESS 41.002 HEAD {CH 2 INTERNAL TRIGGER SENSITIVITY} 41.003 DISP MAKE THE FOLLOWING CONNECTION... 41.003 DISP 41.003 DISP 5520 SCOPE OUTPUT ----------50 Ohm--CH 2 INPUT 41.004 DISP Press SETUP, then press the Default Setup softkey. 41.005 HEAD CH 2 TRIGGER SENSITIVITY: {25 MHz @ 0.35 div}? 41.006 5520 50mVpp 25MH LS S6 S L 41.007 DISP Press AUTOSCALE. 41.008 DISP Set the Volts/Div to 0.1 V/div. 41.009 MESS Slew the amplitude for 0.35 divisions. 41.010 5520 50mVpp 25MH LS S6 N L 41.011 MESS Adjust Trigger Level if needed for a triggered display. 41.011 MESS 41.012 EVAL Will the {display trigger}? 42.001 MESS 42.002 HEAD CH 2 TRIGGER SENSITIVITY: {100 MHz @ 1.0 div}? 42.003 5520 100mVpp 150MH LS S6 S L 42.004 DISP Press AUTOSCALE. 42.005 DISP Set the Volts/Div to 0.1 V/div. 42.006 MESS Slew the amplitude for 1.0 division. 42.007 5520 100mVpp 150MH LS S6 N L 42.008 MESS Adjust Trigger Level if needed for a triggered display. 42.008 MESS 42.009 EVAL Will the {display trigger}? 43.001 MESS 43.002 HEAD {CH 3 INTERNAL TRIGGER SENSITIVITY} 43.003 DISP MAKE THE FOLLOWING CONNECTION... 43.003 DISP 43.003 DISP 5520 SCOPE OUTPUT ----------50 Ohm--CH 3 INPUT 43.004 DISP Press SETUP, then press the Default Setup softkey. 43.005 DISP Select Channel 3 43.006 DISP Set the Volts/Div to 0.1 V/div. 43.007 HEAD CH 3 TRIGGER SENSITIVITY: {25 MHz @ 0.35 div}? 43.008 5520 200mVpp 25MH LS S6 S L 43.009 DISP Press AUTOSCALE. 43.010 MESS Slew the amplitude for 0.35 divisions. 43.011 5520 50mVpp 25MH LS S6 N L 43.012 MESS Adjust Trigger Level if needed for a triggered display. 43.012 MESS 43.013 EVAL Will the {display trigger}? 44.001 MESS 44.002 HEAD CH 3 TRIGGER SENSITIVITY: {100 MHz @ 1.0 div}? 44.003 5520 200mVpp 250MH LS S6 S L 44.004 DISP Press AUTOSCALE. 44.005 MESS Slew the amplitude for 1.0 division. 44.006 5520 100mVpp 250MH LS S6 N L 44.007 MESS Adjust Trigger Level if needed for a triggered display. 44.007 MESS 44.008 EVAL Will the {display trigger}? 45.001 MESS 45.002 HEAD {CH 4 INTERNAL TRIGGER SENSITIVITY} 45.003 DISP MAKE THE FOLLOWING CONNECTION... 45.003 DISP 45.003 DISP 5520 SCOPE OUTPUT ----------50 Ohm--CH 4 INPUT 45.004 DISP Press SETUP, then press the Default Setup softkey. 45.005 DISP Select Channel 4 45.006 DISP Set the Volts/Div to 0.1 V/div. 45.007 HEAD CH 4 TRIGGER SENSITIVITY: {25 MHz @ 0.35 div}? 45.008 5520 200mVpp 25MH LS S6 S L 45.009 DISP Press AUTOSCALE. 45.010 MESS Slew the amplitude for 0.35 divisions. 45.011 5520 50mVpp 25MH LS S6 N L 45.012 MESS Adjust Trigger Level if needed for a triggered display. 45.012 MESS 45.013 EVAL Will the {display trigger}? 46.001 MESS 46.002 HEAD CH 4 TRIGGER SENSITIVITY: {100 MHz @ 1.0 div}? 46.003 5520 200mVpp 250MH LS S6 S L 46.004 DISP Press AUTOSCALE. 46.005 MESS Slew the amplitude for 1.0 division. 46.006 5520 100mVpp 250MH LS S6 N L 46.007 MESS Adjust Trigger Level if needed for a triggered display. 46.007 MESS 46.008 EVAL Will the {display trigger}? 47.001 END